Online Conductivity Controller with 2 set limits and 4-20mA output

Please install and activate YITH WooCommerce Wishlist to add product wishlist
SKU: 160-COND
Download Catalogue

Description

  • Auto/Manual temperature compensation selectable
  • Conductivity and TDS selectable from front panel
  • Cell Factor selectable for both 1.0 and 0.1
  • 4 Step Range selection 0.1µs to 199.9ms for conductivity
  • Digital calibration for Conductivity/TDS
  • Auto Temp Compensation using PT100 sensor
  • Dual set limit with separate relay outputs
  • Settable hysteresis for both set limits
  • Bright Red LED display with separate display for set limit-1
  • Aesthetically designed front panel with membrane switches and ABS cabinet

Frequently Asked Questions — Model 160-COND

What is the 160-COND conductivity controller?

The 160-COND is an online dual-limit conductivity controller by Countronics. Four conductivity ranges (0.1 μS to 199.9 mS), two independent relay outputs, 4-20mA output for SCADA, and auto temperature compensation via PT100.

What does dual-limit mean?

Two independently settable set points each with its own relay. Relay 1 can activate a blowdown valve at a high conductivity limit; Relay 2 can activate dosing at a low limit — automatic two-stage conductivity control.

What is the difference between 160-COND and 5500?

5500 is display-only — no relay or control. 160-COND is a full controller with dual relays and 4-20mA output. Choose 5500 for monitoring only; 160-COND when automatic control is needed.

Does 160-COND have a 4-20mA output?

Yes. 4-20mA output connects to PLCs, SCADA, or data loggers for remote conductivity monitoring alongside local display and relay control.

What industries use the 160-COND?

Boiler feed water, RO plants, cooling tower blowdown, pharmaceutical water systems, food processing, and swimming pool treatment — any process requiring automatic conductivity-based water quality control.

Who manufactures the 160-COND?

Manufactured by Countronics, ISO 9001:2015 certified, established 1978, Ghaziabad. Contact: +91-9810536887.

Additional information

Your custom text or HTML.

Display

0.56” Red LED 7 segment Display for Conductivity/TDS
0.56” Red LED 7 segment Display for Set-Limit-1

Resolution

Conductivity: 0.01ms or 0.01µs depending on range & Cell factor TDS: 1ppm

Auto Temp. Comp

0-100˚C using PT100 external sensor

Manual Temp. Comp

Settable from front panel from 0-100.0˚C

Cond/TDS Calibration

Digital calibration by factor multiplication from 0.75-1.25

Recorder Output

Isolated 4-20mA => 0-2000 (for whatever unit and resolution selected
(Maximum termination 250Ω)

Control Output

Two set of NO-P-NC, 5 Amps potential free Relay Contacts
Relay 1 ON, for Cond/TDS < Set Limit1 – Hysterises1, Relay 1 OFF, for Cond/TDS > Set Limit1
Relay 2 OFF, for Cond/TDS < Set Limit2,Relay 2 ON, for Cond/TDS > Set Limit2 + Hysterises2

Accuracy

0.5% ± 1 Digit

Supply

220V AC ±10% @ 50Hz

Enclosure

Plastic (ABS). Size -96x96x110mmm

Low Voltage

Set at 10.5V internally. For testing 12V Battery AH

Weight

900 grams

Additional Information

Display:
0.56” Red LED 7 segment Display for Conductivity/TDS 0.56” Red LED 7 segment Display for Set-Limit-1
Resolution:
Conductivity: 0.01ms or 0.01µs depending on range & Cell factor TDS: 1ppm
Auto Temp. Comp:
0-100˚C using PT100 external sensor
Manual Temp. Comp:
Settable from front panel from 0-100.0˚C
Cond/TDS Calibration:
Digital calibration by factor multiplication from 0.75-1.25
Recorder Output:
Isolated 4-20mA => 0-2000 (for whatever unit and resolution selected (Maximum termination 250Ω)
Control Output:
Two set of NO-P-NC, 5 Amps potential free Relay Contacts Relay 1 ON, for Cond/TDS < Set Limit1 – Hysterises1, Relay 1 OFF, for Cond/TDS > Set Limit1 Relay 2 OFF, for Cond/TDS < Set Limit2,Relay 2 ON, for Cond/TDS > Set Limit2 + Hysterises2
Accuracy:
0.5% ± 1 Digit
Supply:
220V AC ±10% @ 50Hz
Enclosure:
Plastic (ABS). Size -96x96x110mmm
Low Voltage:
Set at 10.5V internally. For testing 12V Battery AH
Weight:
900 grams