💧 Part of the Water Quality Measurement Hub — guides on pH, ORP, conductivity and TDS.
| Quantity | Common units | Relationship |
|---|---|---|
| Conductivity | µS/cm, mS/cm, S/m | 1 mS/cm = 1000 µS/cm; 1 S/m = 10 mS/cm |
| Resistivity | MΩ·cm, kΩ·cm | MΩ·cm = 1 ÷ (µS/cm) |
| TDS | ppm, mg/L | ppm ≈ µS/cm × factor (0.5–0.7) |
| Salinity | PSU, ppt (‰) | Derived from conductivity and temperature (seawater ≈ 35) |
Pure-water plants (pharma, semiconductor, power) often use resistivity because the numbers are easier to read at very low conductivity.
| Conductivity (µS/cm) | Resistivity (MΩ·cm) | Typical water |
|---|---|---|
| 0.055 | 18.2 | Ultrapure (theoretical limit at 25 °C) |
| 0.1 | 10 | High-purity DI |
| 1.0 | 1.0 | Good DM / pharma purified water |
| 10 | 0.1 | RO permeate (typical) |
| 100 | 0.01 | Soft drinking water |
| Factor | Basis | When used |
|---|---|---|
| 0.5 | Sodium chloride (NaCl) | Many handheld meters; RO and salty water |
| 0.64–0.67 | General / typical | Rough field estimates |
| 0.7 | “442” natural-water mix (sulphate, bicarbonate, chloride) | Drinking water, boilers, cooling towers |
The same water measuring 1000 µS/cm can be reported as 500, 640 or 700 ppm depending on the factor — which is why two meters can “disagree”. Pick one factor for your plant and record it. Background: conductivity vs TDS.
| Water | Conductivity (µS/cm) | TDS approx. (ppm, factor 0.65) |
|---|---|---|
| Ultrapure / DI | < 1 | < 1 |
| RO permeate | 5–50 | 3–30 |
| Drinking water | 50–800 | 30–500 |
| Borewell / hard water | 500–2000 | 300–1300 |
| Cooling tower (concentrated) | 1000–5000 | 650–3250 |
| Seawater | about 50,000 | about 35,000 (salinity) |
Conductivity rises about 2 % per °C for most waters, so readings must be compensated to a reference temperature (normally 25 °C). An uncompensated reading at 35 °C can be 20 % high. All Countronics conductivity instruments use a PT100 sensor for automatic compensation.
Match the cell constant to the range: about K = 0.1 for RO permeate and DM water, K = 1 for drinking and process water, K = 10 for effluent and brine. Details: conductivity cell constant explained.
For measuring and controlling conductivity and TDS:
All models: 220 V AC standard; 110 V AC / 60 Hz, 12 V DC or 24 V DC on request. Test certificates traceable to national laboratories available.
Multiply by a TDS factor, typically 0.5 (NaCl basis) to 0.7 (natural water). For example, 1000 µS/cm ≈ 500–700 ppm.
Resistivity in MΩ·cm = 1 ÷ conductivity in µS/cm. 1 µS/cm equals 1 MΩ·cm.
Theoretically pure water at 25 °C is 0.055 µS/cm, or 18.2 MΩ·cm.
They may use different TDS factors, or one may not be temperature compensated.
1 mS/cm = 1000 µS/cm.