Conductivity cell in water samples on a laboratory bench

Conductivity, Resistivity, TDS And Salinity Units Explained: µS/cm, mS/cm, MΩ·cm And ppm With A Conversion Table

In short: 1 mS/cm = 1000 µS/cm. Resistivity is the inverse: MΩ·cm = 1 ÷ (µS/cm) — ultrapure water at 25 °C is 18.2 MΩ·cm (0.055 µS/cm). TDS is estimated from conductivity: TDS (ppm) ≈ conductivity (µS/cm) × factor, where the factor is usually 0.5 (NaCl basis), 0.64 (general) or 0.7 (442 natural-water basis). Always state the factor and use readings compensated to 25 °C.

💧 Part of the Water Quality Measurement Hub — guides on pH, ORP, conductivity and TDS.

The Units in One Table

QuantityCommon unitsRelationship
ConductivityµS/cm, mS/cm, S/m1 mS/cm = 1000 µS/cm; 1 S/m = 10 mS/cm
ResistivityMΩ·cm, kΩ·cmMΩ·cm = 1 ÷ (µS/cm)
TDSppm, mg/Lppm ≈ µS/cm × factor (0.5–0.7)
SalinityPSU, ppt (‰)Derived from conductivity and temperature (seawater ≈ 35)

Conductivity ↔ Resistivity

Pure-water plants (pharma, semiconductor, power) often use resistivity because the numbers are easier to read at very low conductivity.

Conductivity (µS/cm)Resistivity (MΩ·cm)Typical water
0.05518.2Ultrapure (theoretical limit at 25 °C)
0.110High-purity DI
1.01.0Good DM / pharma purified water
100.1RO permeate (typical)
1000.01Soft drinking water

Conductivity → TDS: Which Factor?

FactorBasisWhen used
0.5Sodium chloride (NaCl)Many handheld meters; RO and salty water
0.64–0.67General / typicalRough field estimates
0.7“442” natural-water mix (sulphate, bicarbonate, chloride)Drinking water, boilers, cooling towers

The same water measuring 1000 µS/cm can be reported as 500, 640 or 700 ppm depending on the factor — which is why two meters can “disagree”. Pick one factor for your plant and record it. Background: conductivity vs TDS.

Typical Values by Water Type

WaterConductivity (µS/cm)TDS approx. (ppm, factor 0.65)
Ultrapure / DI< 1< 1
RO permeate5–503–30
Drinking water50–80030–500
Borewell / hard water500–2000300–1300
Cooling tower (concentrated)1000–5000650–3250
Seawaterabout 50,000about 35,000 (salinity)

Temperature Matters

Conductivity rises about 2 % per °C for most waters, so readings must be compensated to a reference temperature (normally 25 °C). An uncompensated reading at 35 °C can be 20 % high. All Countronics conductivity instruments use a PT100 sensor for automatic compensation.

Choosing the Cell

Match the cell constant to the range: about K = 0.1 for RO permeate and DM water, K = 1 for drinking and process water, K = 10 for effluent and brine. Details: conductivity cell constant explained.

Countronics Instruments for This Job

For measuring and controlling conductivity and TDS:

  • 160-COND — online conductivity controller, two 5 A relays, isolated 4–20 mA, PT100 compensation, 96 × 96 mm panel
  • 160-TDS — online TDS controller, 0–5000 ppm, adjustable TDS factor, two relays, isolated 4–20 mA
  • 5500 — online conductivity indicator with selectable cell factor 0.1 / 1.0, ranges up to 199.9 mS/cm
  • 1004-COND — blind conductivity transmitter (no display), 4–20 mA to PLC, for cell constants 0.1 or 1.0
  • 5501 — bench-top meter for pH, ORP, conductivity, TDS and temperature — for labs and QC

All models: 220 V AC standard; 110 V AC / 60 Hz, 12 V DC or 24 V DC on request. Test certificates traceable to national laboratories available.

Mistakes to Avoid

  1. Comparing TDS values from meters using different factors.
  2. Reading conductivity without temperature compensation.
  3. Using a K = 1 cell on RO permeate.
  4. Mixing up mS/cm and µS/cm (a factor of 1000).

Quick Checklist

  • Units stated (µS/cm or mS/cm)
  • TDS factor stated
  • Compensated to 25 °C
  • Cell constant matches range

Frequently Asked Questions

How do I convert µS/cm to ppm?

Multiply by a TDS factor, typically 0.5 (NaCl basis) to 0.7 (natural water). For example, 1000 µS/cm ≈ 500–700 ppm.

How do I convert conductivity to resistivity?

Resistivity in MΩ·cm = 1 ÷ conductivity in µS/cm. 1 µS/cm equals 1 MΩ·cm.

What is the conductivity of pure water?

Theoretically pure water at 25 °C is 0.055 µS/cm, or 18.2 MΩ·cm.

Why do two TDS meters show different readings?

They may use different TDS factors, or one may not be temperature compensated.

What is 1 mS/cm in µS/cm?

1 mS/cm = 1000 µS/cm.

Need to Measure Conductivity or TDS?

Countronics makes conductivity and TDS controllers, indicators and transmitters. Tell us your water and range and we will suggest a model and cell.


In short: Most industrial water systems are controlled with four online measurements: pH (corrosion, scaling, chemistry), ORP (disinfection and oxidation), conductivity/TDS (dissolved salts, purity, blowdown) and temperature (compensation and process). The table below
Water Quality Parameters By Industrial Application: What To Monitor In RO Plants, Boilers, Cooling Towers, ETPs And Plating Rinses
In short: Use an online pH controller or indicator when pH must be watched or controlled continuously in a tank or pipe — it has relays for dosing and a 4–20 mA output.
Online vs Portable vs Bench-Top pH Meters: Which One Do You Need For Process, Field And Lab?
In short: 1 mS/cm = 1000 µS/cm. Resistivity is the inverse: MΩ·cm = 1 ÷ (µS/cm) — ultrapure water at 25 °C is 18.2 MΩ·cm (0.055 µS/cm). TDS is estimated from conductivity: TDS
Conductivity, Resistivity, TDS And Salinity Units Explained: µS/cm, mS/cm, MΩ·cm And ppm With A Conversion Table