An ORP electrode measures the voltage between a platinum (or gold) sensing surface and a reference electrode. Oxidisers such as chlorine, ozone and hypochlorite push the reading up; reducing agents such as sodium bisulphite or metabisulphite pull it down. ORP does not tell you the exact concentration of a chemical, but it tracks the activity of the reaction — which is what actually matters for disinfection and effluent treatment. For the basics, see what is ORP.
| Application | Typical ORP | Notes |
|---|---|---|
| Swimming pool disinfection | +650 to +750 mV | Keep pH at 7.2–7.6 or the chlorine is far less effective |
| Cooling tower / drinking water disinfection | +500 to +700 mV | Depends on oxidiser and contact time |
| Cyanide oxidation (alkaline chlorination) | Above about +300 mV, pH 10 or higher | Never allow pH to fall — toxic gas risk |
| Hexavalent chrome reduction | About +200 to +300 mV at pH 2–3 | ORP falls as Cr⁶⁺ is reduced to Cr³⁺ |
| Dechlorination before RO | Low positive mV | Protects RO membranes from free chlorine |
These are starting points only. Every plant should confirm its set point with lab tests, then fine-tune the controller.
ORP changes with pH. In a pool, the same chlorine dose gives a much lower ORP at pH 8 than at pH 7.2, because less of it is in the active hypochlorous acid form. In chrome and cyanide treatment, the chemistry only works inside a narrow pH window. That is why ORP systems are almost always paired with pH control — see the 7215-01 pH and ORP controller, which handles both.
| Problem | Likely cause | Fix |
|---|---|---|
| Slow, sluggish response | Coated platinum surface | Clean gently; condition in fresh tap water |
| Reading drifts over days | Reference junction clogged | Clean junction or replace electrode |
| Erratic jumps | Air bubbles or poor earthing | Mount at an angle, check earthing and cable |
| Wrong value in check solution | Aged electrode | Replace — typical life 1–2 years |
Most pools are controlled at about +650 to +750 mV with pH held between 7.2 and 7.6. Check local health regulations for the exact requirement.
Not directly. ORP shows how strongly the water is oxidising, which depends on chlorine, pH, temperature and other substances. Use a chlorine test kit to relate the two at your site.
Usually the platinum surface is coated or the reference junction is clogged. Clean the electrode gently and, if it is more than one to two years old, replace it.
In almost every application, yes. ORP depends on pH, and both cyanide oxidation and chrome reduction only work in a defined pH range. A combined pH/ORP controller is a convenient option.
Check it monthly in a standard ORP test solution and clean it when the response becomes slow.
Countronics makes pH and ORP indicators and controllers with relay limits and 4–20 mA outputs. Share your application and we will suggest set points and a suitable model.