A pH glass bulb works only when its surface layer is hydrated. Storing it dry ruins the response; storing it in distilled water slowly leaches the reference electrolyte. Use the storage cap filled with 3M KCl solution, or pH 4 buffer if KCl is not available.
| Fouling | Cleaning method |
|---|---|
| Oil, grease | Mild detergent solution, then rinse |
| Scale (calcium, carbonates) | Dilute hydrochloric acid (about 0.1 M) for a few minutes |
| Protein (dairy, food, pharma) | Pepsin / enzyme cleaning solution |
| Sulphide (black junction) | Special sulphide cleaner or replace |
| General dirt | Rinse; wipe gently with soft tissue — do not rub the bulb |
After any chemical cleaning, soak in storage solution for at least 30 minutes and recalibrate.
| Check | Good | Replace / service |
|---|---|---|
| Offset at pH 7 | Within about ±30 mV | Beyond about ±30–40 mV |
| Slope | 95–102% | Below about 90% |
| Response time | Stable within ~30 s | Slow, drifting |
Most controllers and meters show offset and slope after a two-point calibration. Record them each time — a falling trend tells you when to order a spare.
In clean water an electrode may last one to two years; in hot, dirty or high-pH effluent it may last only months. Keeping a spare avoids downtime. For selection tips, see the pH electrode selection guide.
Wet, in 3M KCl storage solution. Never store it dry or in distilled water.
Soak briefly in mild detergent, rinse, then soak in storage solution and recalibrate.
When the slope falls below about 90%, the offset is large, or response becomes slow even after cleaning.
Typically one to two years in clean applications, less in hot or dirty process streams.
Often a clogged reference junction, a dirty bulb or old buffers. Clean, recalibrate and check slope and offset.